Authors
Swapan K Chowdhury, Viswanatham Katta, Brian T Chait
Publication date
1990/3
Journal
Rapid Communications in Mass Spectrometry
Volume
4
Issue
3
Pages
81-87
Publisher
Heyden & Son Limited
Description
The construction and performance of an electrospray‐ionization mass spectrometer with new features are described. The mass spectrometer consists of a newly designed electrospray ion‐source that is plugged directly into a modified commercial quadrupole mass spectrometer with the ions entering the mass analyzer through a long metal capillary tube and three stages of differential pumping. The present ion source differs from previous designs in the combination of techniques employed in the transportation and desolvation of solvated biomolecule ions, prior to mass analysis. Transport of ionized entities between atmospheric pressure and vacuum is carried out through a 203 mm long stainless steel capillary tube with a 0.5 mm bore. Desolvation is effected by the use of controlled heat transfer through the long capillary tube and collisional activation in a region of reduced pressure between the capillary tube exit …